Omega 3200ST Manuale Utente

Pagina di 26
M-4271/0707, pg. 17 of 26 
7.  Zero Adjustments 
 
The zero should be checked as detailed in section C part 2. If an 
adjustment is needed the Zero Potentiometer should be carefully turned 
until the output (VDC) becomes zero.  
 
 
Caution: Do NOT adjust the Gain Potentiometer when adjusting 
the zero or the unit will need to be recalibrated. 
 
 
 
Making Zero Adjustments Using a Small Flathead Screwdriver 
 
 
 
 
Care should be taken to only make small adjustments to the zero 
potentiometer. If too much of an adjustment is made and difficulties are 
being experienced in achieving a zero reading then turn the potentiometer 
fully anti-clockwise and begin making small clockwise adjustments until a 
zero reading is obtained. 
 
 
8.  Recalibration 
 
If recalibration is required please contact the Omega Engineering 
Customer Service Department.  
 
 
9.  Changing the Calibration Gas – FMA 3400/3400ST Series Only 
 
The FMA 3400/3400ST Series may be calibrated for up to three gases. 
These gases, their corresponding flow ranges and accuracy specifications 
are detailed on the calibration certificate.  
 
The calibration gas required is determined by selecting the corresponding 
dip switch on the front panel of the display.