Omega CL526 Manuale Utente

Pagina di 50
35
10
DIGITAL INTERFACE
The  CL526 portable  calibrator is equipped with a digital interface. The interface circuit is essentially based on the serial
communication interface subsystem (SCI) on the chip of the microprocessor. The output voltage levels are TTL at 0 to +5
V.
An optional adaptor to convert the voltage level from 0 to +5V to RS232 levels can be supplied on request. This adaptor
is  required to interface the CL526 with a Personal Computer.
10.1
Digital interface data program mode
  To enter the procedure, press  the <ENTER> + <ON> keys. The display will indicate:
CAL?  65388   N=0  
  To enter the program mode press the <2> key;
Baud Rate   19200  
The numerical value of the “baud rate” can be one of the following : 19200, 9600, 4800, 2400, 1200, 600, 300
  Select, with the <
>
 or <
>
 key the “baud rate” used by the receiver unit and transmission lines.
 Press 
the 
 
<2>
 key to memory load the baud rate.
  The display will indicate:
ID-Name :      1  
The number represents the address code assigned to the instrument.
  Press the <▲> or <▼> key to select a number from 00 to 99.
 Press 
the 
 
<2>
 key to memory load the programmed value.
  The display will return to the original indication:
CAL?  65388   N=0   
  To exit the procedure press the <OFF> key.
10.2
Digital output wiring practice
The wiring to the digital output signals is made through a mini DIN connector mounted on the lower end of the case.
The pertinent connections are indicated below.
ground
5 V 
Tx
Rx
Female miniDIN connector  
(case mounted - external view)
ground
For easy interconnections a miniDIN connector with cable (cat. EE420123) con be supplied on request. The conductors
color codes can change with different supplier; please check before using.