Installing the System 
 
 
Part No. 93463 / Drawing No. 0921-01561 
Page 32 of 221 
Revision A (Sep 2007) 
Alternatively, the analog port can be configured to output QF on a 0–1 V scale.  
The QF can then be monitored by connecting a panel meter to the analog 
output and displaying the voltage.  A third method of monitoring the Quality 
Factor is to reassign one of the pulse outputs to output the Quality Factor as a 
frequency output. 
To ensure quality measurements, the laser beams must maintain “contact” 
with the product.  If the product exhibits large movements during operation, it 
should be stabilized with external roller guides.  Beta LaserMike offers optional 
product guides.  Contact your local Beta LaserMike sales representative or the 
factory for additional information. 
Material Present Adjustment 
The intensity of the light reflected from the measurement area can be used to 
determine whether material is in the measurement zone or not. To 
compensate for varying reflectivity of different materials, it is often necessary 
to adjust the threshold of the light detection circuit. This adjustment can be 
made with a serial command or by using the LaserTrak software. 
Material Present Dropout Time Selection 
If you are taking measurements and the Material Present signal is lost for less 
than the selected Material Present dropout time, the loss is considered to be a 
momentary loss of signal and is ignored.  If the signal is lost for longer than the 
selected Material Present dropout time, it is considered the end of the material.  
This function is very useful when making length measurements on poor 
reflecting material. 
Validation Time Limit Selection 
The validation time limit is used to start making length calculations.  Two 
conditions are necessary to start a length measurement: Material Present and 
valid velocity data.  When material enters the measurement volume, the 
system determines the time between the Material Present signal and the first 
valid data.  If the time is less than or equal to the validation time limit 
(selectable via LaserTrak or ‘G’ command), then the start-up length-correction 
is calculated and a length measurement is initialized.  If the time between the 
Material Present signal and the first valid velocity data is longer than the 
validation limit, then the length calculation is aborted and an error or fault is 
sent.  This error is sent by setting the length output (either in continuous output 
mode or Final Length Mode) to all 9’s.