Emerson NGA 2000 Manuale Utente
NGA-2000 System Calibration
Supplement - 12
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
A further general parameter is "Timeout for Gastest". It is located in the menu "Test
procedures".
procedures".
SCTESTGAS
SCTIMEOUT
SCTIMEOUT
SCCALTIME1
SCSTEPMOD
SCSTEPMOD
This parameter is related to the possibility to activate a gas valve of a specific module for test
purposes. Here we can determine a time after which an activated gastest automatically
switches back onto SAMPLE-Gas.
Is this parameter set to "0" no automatic back switching is done.
purposes. Here we can determine a time after which an activated gastest automatically
switches back onto SAMPLE-Gas.
Is this parameter set to "0" no automatic back switching is done.
System configuration and diagnostics...
↓
System calibration...
↓
Calibration/Test Procedures...
↓
Test procedures...
↓
-- Test Procedures --
Gastest of specific module SAMPLE-Gas
Timeout for Gastest (0 = no timeout) 180 s
Choose specific analyzer module...
Timeout for Gastest (0 = no timeout) 180 s
Choose specific analyzer module...
_______________________________________________
Procedure time: 10 s
Analyzer-module:
Procedure time: 10 s
Analyzer-module:
TFID