Emerson NGA 2000 Manuale Utente
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 11
Line of variables "Span ranges":
♦
Together:
All measurement ranges of the same channel will be calibrated together.
This is the usual calibration mode.
All measurement ranges of the same channel will be calibrated together.
This is the usual calibration mode.
♦
Separately:
Each measurement range will be calibrated separately.
Each measurement range will be calibrated separately.
Line of variables "Valve position":
Options:
♦
Samplegas
♦
Zerogas
♦
Purgegas
♦
Testgas
♦
Spangas-1...-4
♦
Linearizer
♦
Spangas
♦
AllClosed
♦
Other-Proc.
♦
Basic-Status
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Calibration procedure setup