Emerson NGA2000 Manuale Utente

Pagina di 178
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
5 - 55
Main Menu — Expert controls and set up
↓↓
Expert analyzer controls
↓↓
Range settings
↓↓
In the menu "Range Settings" you can set up the begin and the end of each range. These
values will only be limited by the setups in the line "Minimum range" resp. "Maximum
range". The concentrations of these lines are manufacturer's settings and cannot be
changed by the customer.
Set up parameters:
♦ 
Select any line of variables with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable with the ENTER key or the 
 -key.
♦ 
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key
resp. select the whole parameter with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
HOME
ESCAPE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Range Settings
Minimum range:
10.0 ppm
Range 1 lower limit:
0.0 ppm
Range 1 upper limit:
10.0 ppm
Range 2 lower limit:
0.0 ppm
Range 2 upper limit:
25.0 ppm
Range 3 lower limit:
0.0 ppm
Range 4 lower limit:
0.0 ppm
Range 4 upper limit:
250.0 ppm
Maximum range:
10000 ppm
Range 3 upper limit:
100.0 ppm
5.5.1 Expert Analyzer Controls
Range Settings