Saving the factors:
♦ 
You can store the actual zero offset and span factor with the F2 key (STORE) of the
menu "Range 1/2/3/4 Factors". You will find the stored settings on the second menu
page (see next illustration).
♦ 
You can change to the second menu page of the menu "Range 1/2/3/4 Factors" with
the F4 key (HISTORY):
♦ 
You can change the stored settings against the manufacturer's settings with the F3 key
(RSTR MN: Restore the manufacturer's values). The "Zero offset" and the "Span factor"
of the first menu page will change according to the setups of the manufacturer's
settings.
♦ 
You can change the manufacturer's settings against the stored settings with the F4 key
(RSTR ST: Restore the "stored" values). The "Zero offset" and the "Span factor" of the
first menu page will change according to the setups of the stored settings.