Emerson NGA2000 Manuale Utente

Pagina di 178
6 - 26
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Technical level configuration
↓↓
Diagnostic menus
↓↓
Control module diagnostics
↓↓
In the menu "Control Unit Diagnostics" you can find software error messages of the
control module. If such messages are available, you can act in the following way:
1) Make a note of the error messages.
2) Reset the message:
• 
Press the ENTER key or the 
 -key in the line "Edit to reset".
• 
Set up "Reset" with the 
↑↑
 -key or the 
↓↓
 -key and confirm it with the ENTER key:
The error messages will disappear if the reason is not existing any longer.
• 
If the error messages keep displayed:
3) System reset:
• 
Change to the menu "System Reset" with the F3 key (REBOOT).
• 
Press the ENTER key or the 
6-14).
4) Check the error messages:
• 
Go again to the menu "Control Unit Diagnostics".
• 
If the error messages are still existing, contact your customer service, please.
• 
If the error messages are disappeared, you should set up the report status again:
Set up "Report" in the line "Edit to reset".
HOME
ESCAPE
REBOOT
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Control Unit Diagnostics
Software error code (1 = no error):
1
And:
And:
And:
And:
And:
And:
Edit to reset:
Report
Last software error message
And:
6.3.1 Diagnostic Menus
Control Module Diagnostics