Main Menu — Technical level configuration
↓↓
Diagnostic menus
↓↓
Analyzer module diagnostics
↓↓
Start up analyzer
↓↓
In the menu "Analyzer start up" you will find the actual and target temperatures of the
converter and of the detector block. Besides, you can reboot the analyzer module and re-
initialize the EPROM in this menu:
1) Reboot:
• 
To restart the analyzer you have to press the F3 key.
• 
This reboot will start your analyzer in the same way as you do by switching on.
• 
If you flow the analyzer with zero gas during this time: The NO signal should be "0".
2) Re-initializing of the EPROM:
• 
If you press the F4 key you will update the EPROM data.
• 
Caution:
The analyzer will go back to the default values using this function !
After that you have to do a new start up, completely !
HOME
ESCAPE
REBOOT
INIT.
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Analyzer start up
(Press INIT to re-initialize the EEPROM data)
Sensor temperature:
27.0 C
Detector block temperature:
50.3 C
Target temperature:
51.5 C
Converter temperature:
302.0 C
Target temperature:
300.0 C
Analyzer state:
STANDBY
6.3.2.9 Diagnostic Menus
Analyzer start up