Emerson NGA2000 Manuale Utente

Pagina di 178
4 Basic Controls
90003479(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 1
In the chapter "Basic Controls" you will find the most important functions to set up your
CLD analyzer module via an NGA front panel:
1) Measurement  p. 4 - 3pp  (Chapter 4.1.1 and 4.1.2, p. 4-3 to 4-6, are only important 
for systems with several channels.)
All steps are figured with detailed illustrations and operation instructions. In the left column
you can see display and keyboard of the NGA front panel. In the right column you can
read the instructions and notes. All instructions will begin with any single component
display and will end with the corresponding single component display after the setups are
done. The keys you have to press are illustrated in black. The instructions are completed
with notes and further informations.
Example: You want to change from the single component display of NO
x
 to the main menu
• 
Picture one shows the starting situation: single component display of NO
x
.
• 
If you press the F3 key you will change to menu page shown in picture 2: Main Menu.
Left column:
Right column:
LCD and keyboard
Instructions and notes
DISPLAY
STATUS
MENU
NEXT
INFO
CLD
Sample press.:
Detector temperature:
Block temperature:
Converter temp.:
51.5 C
0.5 C
340 hPa
360 C
F1
F2
F3
F4
F5
0
250
95.00
ppm NOx
Range: 4
MEASURE
STATUS
NEXT
LOCK
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Main Menu
Basic controls...
Expert controls and set up...
Technical level configuration...
(Operational configuration)
(Diagnostic and manufacturing/service)
⇒ 
⇒ 
Change to the
Main Menu
Press  
F3
or  
⇒ 
⇒ 
Next instruction
or step,
here e.g.:
Change to the
Basic Controls:
Press  
or