Emerson NGA2000 Manuale Utente
System Calibration
Supplement - 16
NGA 2000
90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98
Here we first choose the specific analyzer module and then we select at the parameter
"Gastest of specific module" the required gas.
"Gastest of specific module" the required gas.
Each of the following types is possible:
•
SAMPLE-GAS (test mode deactivated)
•
ZERO-GAS
•
SPANGAS-1 (spangas for range 1)
•
SPANGAS-2 (spangas for range 2)
•
SPANGAS-3 (spangas for range 3)
•
SPANGAS-4 (spangas for range 4)
With parameter "Timeout for Gastest" we can determine a time after which the activated
gastest switches automatically back onto SAMPLE-GAS.
gastest switches automatically back onto SAMPLE-GAS.
Is this parameter set to "0" the automatic back switching is not done automatically and
user must break this mode with a "Cancel Calibration"-Command.
user must break this mode with a "Cancel Calibration"-Command.