Emerson NGA2000 Manuale Utente

Pagina di 178
4.1.2  Measurement - 
Multi Component Display:
Sequence setup
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
NGA 2000
4 - 5
DISPLAY
STATUS
MENU
NEXT
INFO
CLD
Sample press.:
Detector temperature:
Block temperature:
Converter temp.:
51.5 C
0.5 C
340 hPa
360 C
F1
F2
F3
F4
F5
0
250
95.00
ppm NOx
Range: 4
INFO
CLD
F1
F2
F3
F4
F5
SELECT
STATUS
TAGS OFF
MLT25/CH2/R2
ppm SO2
333.0
MLT25/CH3/R2
ppm NO
150.0
MLT25/CH4/R2
%O2
20.00
CLD
ppm NOx
95.00
MLT25/CH1/R2
ppm CO
45.00
INFO
CLD
F1
F2
F3
F4
F5
SELECT
STATUS
TAGS OFF
MLT25/CH2/R2
ppm SO2
333.0
MLT25/CH3/R2
ppm NO
150.0
MLT25/CH4/R2
%O2
20.00
CLD
ppm NOx
95.00
MLT25/CH1/R2
ppm CO
45.00
⇒ 
⇒ 
Change to the multi
component display
Press  
F1
⇒ 
⇒ 
Enable the selecting:
">"
Press  
F1
or  
Note:
If no channels of other
analyzer modules are
combined to the CLD,
only the bargraph of the
CLD module will be displayed.
Then, the sequence setup
will not be possible !
⇒ 
⇒ 
Select the channel
for position one
Press  
  or  
as often as necessary to
put the > - mark into the
selected line.
Example:
MLT channel 2 (SO
2
) shall
appear at position one of the
multi component display.