5 - 30
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Expert controls and set up
↓↓
Auxiliary module controls
↓↓
Set-up SIO module
↓↓
From the menu "LOCAL SIO CONFIGURATION SECTIONS" you can change to
submenus to control and set up several output configurations of the auxiliary SIO module.
General Configuration of a SIO board (the whole specification is in another manual):
♦ 
Analog outputs: minimum 2, maximum 8
♦ 
Serial interface (RS 232 or RS 485) to connect the analyzer with a external computer
♦ 
Three relay outputs
If there is a SIO module installed in the platform or in the MLT analyzer, you have to put in
"Yes" in the line "Module installed", otherwise "No".
Set up parameters:
♦ 
Select the line of variables or any menu line with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable or change to the submenu with the ENTER key or the 
 -key.
♦ 
Select the parameter with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the ENTER key or cancel and go back to the last value with
the F2 key.
EXPERT
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
- LOCAL SIO CONFIGURATION SECTIONS -
Analog output setup...
Serial interface setup...
Configuration of Relay Outputs...
Module installed:
Yes
5.2.1 Auxiliary Module Controls
Local SIO Configuration Sections