Omega OMB-DAQBOARD-3000 Manuale Utente

Pagina di 108
DaqCal                                                         
User Calibration Utility
 
 
Contents 
 
CAUTION
 
 
 
 
 
Turn off power to all devices connected to the system before connecting cables or 
setting configuration jumpers and switches.  Electrical shock or damage to 
equipment can result even under low-voltage conditions. 
 
CAUTION
 
 
The discharge of static electricity can damage some electronic components.  
Semiconductor devices are especially susceptible to ESD damage.  You should 
always handle components carefully, and you should never touch connector pins or 
circuit components unless you are following ESD guidelines in an appropriate ESD 
controlled area.  Such guidelines include the use of properly grounded mats and 
wrist straps, ESD bags and cartons, and related procedures. 
 
Overview 
DaqCal is a Windows-based program used to calibrate Daq systems, for example: DaqBoards, DaqBooks, 
DaqScans, and DaqLabs.  DaqCal can also be used to calibrate optional analog expansion cards. 
Daq devices are factory-calibrated and you do not need to recalibrate the devices upon initially receiving 
them. 
Note:  Certain DBK options are self-calibrating and do not require manual calibration by the user.  Such 
devices are disassociated from DaqCal, i.e., you cannot use DaqCal to calibrate them. 
User Calibration Guide, 457-0931 rev 1.0 
918494
 DaqCal_pg.