A&D AD-4212B-101/102/201/301 Manuale Utente

Pagina di 102
 
54
14-2  Memory for Weighing Data 
  The data memory function can store 200 sets of weighing data (100 set if time and date are 
added). Even if the AC adapter is removed, the data is maintained in non-volatile memory. 
 It is not necessary that the printer or computer be continually connected to the balance, 
because the balance stores the weighing data in memory. 
  The data in memory is available to be displayed on the balance for confirmation, or to output 
several sets of data at one time to a printer or personal computer. In the function table, what 
data is to be added to the output data (ID number, data number, time and date) can be 
selected. 
 
Storing the weighing data 
Note 
If “
Clr” appears blinking in the upper left of the display, delete the data in memory. 
1  Set the “Data memory ( data )” parameter to “2”. 
2  Set the “Time/Date output (5-td)” parameter as necessary. 
3  The storing operation depends on the “Data output mode ( prt )” parameter setting. Four types 
of operating modes are available to store data. 
Key mode 
Each time the PRINT key is pressed while the displayed value is 
stable, the balance stores the weighing data. 
Auto print modes A and B 
When  the  displayed  value is stable and the conditions of “Auto 
print polarity”, “Auto print difference” and reference value are met, 
the balance stores the weighing data.   
Interval memory mode 
Weighing data is stored at an interval specified in “Interval time 
(int)”. Press the PRINT key to start and stop this mode. 
(Display example: AD-4212A series balance) 
 
 
 
 
 
 
 
Caution 
  When weighing data is being stored in memory, the data output to a personal computer using 
the RS-232C interface is not available. 
    “fUl” indicates that memory is full or the memory capacity has been reached. More data can 
not be stored unless the memory data is deleted.