A&D AD-4212A/B Manuale Utente

Pagina di 102
 
43
 
 
Five-level comparison 
Contact output 
Comparison 
results 
Display 
H H 
HI OK  LO LL 
Buzzer
H H 
HH  blinking 
short
open
open
open open 
   
HI 
HI illuminated 
open
short
open
open open 
     
OK 
OK illuminated 
open
open
short
open open 
 
LO 
LO illuminated 
open
open
open
short open 
 
LL 
LL blinking 
open
open
open
open short 
     
* Set the upper limit value equal to or higher than the lower limit value. 
Set the secondary upper limit value equal to or higher than the upper limit value. 
Set the secondary lower limit value equal to or lower than the lower limit value. 
 
Setting the upper/lower limit values 
When five-level comparison is to be performed, using the AD-4212B series balance, set the 
“Comparator output setting ( Cp-t )” parameter to “1”. 
 
1  Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.
 
 
2  Press the SAMPLE key several times to display the value name to be set. 
     CpPHH  secondary upper limit value 
     CpPHi  upper limit value   
     CpPlo  lower limit value 
     CpPll  secondary lower limit value   
3  Press the PRINT key. The value currently set is displayed with all of the digits blinking. 
4  When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 5. 
  When the current setting is to be changed, press the RE-ZERO key. Change the setting using 
the following keys. 
SAMPLE key 
To select the digit to change the value. 
RE-ZERO key 
To change the value of the digit selected.   
MODE key 
To switch the polarity. 
PRINT key 
To store the new setting and go to step 5. 
CAL key 
To cancel the new setting and go to step 5. 
5  Repeat steps 2-4 to set values as necessary. 
Note 
The upper/lower limit values can be set by external commands using the RS-232C serial interface. 
Command example 
H H :+100.00000 
  g  (to set secondary upper limit value) 
 H I  
:+080.00000 
(to 
set 
upper limit value) 
 
L O  :+060.00000 
g  (to set lower limit value) 
 LL 
:+040.00000 
(to 
set 
secondary lower limit value) 
 
  
       indicates a space (20h)