81
 
 
 Display             Error code 
Description 
EC, E03 
Timeout error 
If the timeout parameter is set to “t-Up1”, the balance 
did not receive the next character of a command within 
the time limit of one second. 
Confirm the communication. 
EC, E04 
Excess characters error 
The balance received excessive characters in a 
command. 
Confirm the command. 
EC, E06 
Format error 
A command includes incorrect data. 
e.g. The data is numerically incorrect. 
Confirm the command. 
EC, E07 
Parameter setting error 
The received data exceeds the range that the balance 
can accept.   
Confirm the parameter range of the command. 
Other errors 
If the errors described above can not be released or 
other errors are displayed, contact the local A&D dealer. 
 
22-3  Asking For Repair 
If the balance needs service or repair, contact the local A&D dealer. 
The balance is a precision instrument. Use much care when handling the balance and observe the 
following when transporting the balance. 
  Use the original packing material. 
  Remove the weighing pan from the weighing unit.