RS232 Interface 
RS232 Interface Connector 
The 9-way D-type serial interface connector is located on the instrument rear panel. The pin 
connections are as shown below: 
Pin Name  Description 
No internal connection 
TXD 
Transmitted data from instrument 
RXD 
Received data to instrument 
No internal connection 
5 GND 
Signal 
ground 
No internal connection 
RXD2 
Secondary received data (addressable RS232 only) 
TXD2 
Secondary transmitted data (addressable RS232 only)  
GND 
Signal ground (addressable RS232 only) 
Single Instrument RS232 Connections 
For single instrument remote control only pins 2, 3 and 5 are connected to the PC. However, for 
correct operation links must be made in the connector at the PC end between pins 1, 4 and 6 and 
between pins 7 and 8, see diagram. Pins 7 and 8 of the instrument must not be connected to the 
PC, i.e. do not use a fully wired 9–way cable. 
 
Baud Rate is set as described above in Address and Baud Rate Selection; the other parameters 
are fixed as follows: 
Start Bits:  1 
Parity:  None 
Data Bits:  8 
Stop Bits:  1 
 Addressable RS232 Connections 
For addressable RS232 operation pins 7, 8 and 9 of the instrument connector are also used. 
Using a simple cable assembly, a 'daisy chain' connection system between any number of 
instruments, up to the maximum of 32 can be made, as shown below: 
 
The daisy chain consists of the transmit data (TXD), receive date (RXD) and signal ground lines 
only. There are no control/handshake lines. This makes XON/XOFF protocol essential and allows 
the inter-connection between instruments to contain just 3 wires. The wiring of the adaptor cable 
is shown below: 
 28