Remote Commands 
RS232 Remote Command Formats 
Serial input to the instrument is buffered in a 256 byte input queue which is filled, under interrupt, 
in a manner transparent to all other instrument operations. The instrument will send XOFF when 
approximately 200 characters are in the queue. XON will be sent when approximately 100 free 
spaces become available in the queue after XOFF was sent. This queue contains raw (un-
parsed) data which is taken, by the parser, as required. Commands (and queries) are executed in 
order and the parser will not start a new command until any previous command or query is 
complete. In non–addressable RS232 mode responses to commands or queries are sent 
immediately; there is no output queue. In addressable mode the response formatter will wait 
indefinitely if necessary, until the instrument is addressed to talk and the complete response 
message has been sent, before the parser is allowed to start the next command in the input 
queue. 
Commands must be sent as specified in the commands list and must be terminated with the 
command terminator code 0AH (Line Feed, LF). Commands may be sent in groups with 
individual commands separated from each other by the code 3BH (;). The group must be 
terminated with command terminator 0AH (Line Feed, LF). 
Responses from the instrument to the controller are sent as specified in the commands list. Each 
response is terminated by 0DH (Carriage Return, CR) followed by 0AH (Line Feed, LF). 
<WHITE SPACE>
 is defined as character codes 00H to 20H inclusive with the exception of those 
which are specified as addressable RS232 control codes. 
<WHITE SPACE>
 is ignored except in command identifiers. e.g. '*C LS' is not equivalent to '*CLS'.  
The high bit of all characters is ignored. 
The commands are case insensitive. 
GPIB Remote Command Formats 
GPIB input to the instrument is buffered in a 256 byte input queue which is filled, under interrupt, 
in a manner transparent to all other instrument operations. The queue contains raw (un-parsed) 
data which is taken, by the parser, as required. Commands (and queries) are executed in order 
and the parser will not start a new command until any previous command or query is complete. 
There is no output queue which means that the response formatter will wait, indefinitely if 
necessary, until the instrument is addressed to talk and the complete response message has 
been sent, before the parser is allowed to start the next command in the input queue. 
Commands are sent as 
<PROGRAM MESSAGES>
 by the controller, each message consisting of zero 
or more 
<PROGRAM MESSAGE UNIT>
 elements separated by 
<PROGRAM MESSAGE UNIT SEPARATOR>
 
elements. 
<PROGRAM MESSAGE UNIT>
 is any of the commands in the remote commands list. 
<PROGRAM MESSAGE UNIT SEPARATOR>
 is the semi-colon character ';' (3BH). 
<PROGRAM MESSAGES>
 are separated by 
<PROGRAM MESSAGE TERMINATOR>
 elements which may 
be any of the following: 
 
NL 
 
The new line character (0AH)  
 
NL^END 
The new line character with the END message 
 
^END   
The END message with the last character of the message 
Responses from the instrument to the controller are sent as 
<RESPONSE MESSAGES>
. A 
<RESPONSE MESSAGE>
 consists of one 
<RESPONSE MESSAGE UNIT>
 followed by a 
<RESPONSE 
MESSAGE TERMINATOR>
<RESPONSE MESSAGE TERMINATOR>
 is the new line character with the END message NL^END. 
Each query produces a specific 
<RESPONSE MESSAGE>
 which is listed along with the command in 
the remote commands list. 
 36