Calibration 
Routine calibration is carried out without opening the instrument. 
Allow 10 minute warm-up before commencing calibration. 
Equipment Required 
A 5.5 digit multimeter with better than 0.02% accuracy on dc volts and better than 0.12% 
accuracy on dc current (to 5A); alternatively use a precision shunt for current measurement. 
Calibration 
Assign CONTROL to Output 1. To enter calibration press SHIFT, #, 99.  If the instrument is a long 
way out of calibration, the default calibration values should be loaded first by pressing SHIFT, #, 
91.  Pressing ESCAPE at anytime will abort the calibration procedure and revert to the stored 
calibration values. 
Having entered calibration mode, follow the table below. To increment to the next step press OK. 
The calibration must be done in sequence. At the end of calibration press OK to store the 
calibration. The instrument automatically sets the range and settings at each step. 
Use the spin-wheel to adjust calibration; holding down the  JOG SET OFF  key will give x100 
increments to speed up calibration. Pressing  SHIFT will show the calibration value.   
Repeat for Output 2. 
 
 
 
Step/Range Settings 
Adjust 
for:- 
Load 
Detail 
1, 35V 3A 
0.01V 3A 
10mV ± 0.5mV 
DVM 
o/p 0V 
2, 35V 3A 
0.01V 3A 
0.010V on display 
DVM 
V readback zero 
3, 35V 3A 
35V 3A 
35V ± 0.5mV 
DVM 
o/p V span 
4, 35V 3A 
35V 3A 
35.000 on display 
DVM 
V readback span 
5, 35V 3A 
0.01V 3A 
10mV ± 0.5mV 
DVM 
o/p 0V 
6, 35V 3A 
35V 3A 
35V ± 0.5mV 
DVM 
o/p V span 
7, 35V 3A 
2V 0.001A 
1mA ± 0.5mA 
milli-ammeter 
o/p A offset 
8, 35V 3A 
2V 0.001A 
0.001 on display 
milli-ammeter 
A readback zero 
9, 35V 500mA 
2V 0.1mA 
0.1mA ± 0.05mA 
milli-ammeter 
o/p mA offset 
10, 35V 500mA 
2V 0.1mA 
flashing 0.0/0.1 on display 
milli-ammeter 
mA readback zero 
  
 change 
load 
 
11, 15V 5A 
2V 4A 
4A ± 0.5mA 
ammeter 
o/p A span 
12, 15V 5A 
2V 4A 
4.000 on display 
ammeter 
A readback span 
13, 15V 5A 
2V 4.1A 
4.100 on display 
ammeter 
A readback 13bit 
14, 35V 500mA 
2V 400mA 
400mA ± 0.05mA 
ammeter 
o/p mA span 
15, 35V 500mA 
2V 400mA 
400.0 on display 
ammeter 
mA readback span 
16, 35V 500mA 
2V 410mA 
410.0 on display 
ammeter 