11-2      Specifications 
898095
 DaqScan/2000 
Series 
 
 
Ranges:  Software programmable via sequencer on a per-channel basis. 
Accuracy (Note 2)  
One Year, 0 to 35°C 
±(% reading + % range) 
Input Noise (Note 3)  
(LSB rms) 
10 Hz to 200 kHz 
Voltage Range 
(Note 1) 
Absolute  
Typical  
0 to +10 V 
0.015 + 0.005 
0 to +5 V 
0.015 + 0.005 
0 to +2.5 V 
0.015 + 0.005 
0 to +1.25 V 
0.015 + 0.008 
0 to +0.625 V 
0.015 + 0.008 
0 to +0.3125 V 
0.015 + 0.008 
-10 to +10 V 
0.015 + 0.005 
-5 to +5 V 
0.015 + 0.005 
-2.5 to +2.5 V 
0.015 + 0.005 
-1.25 to +1.25 V 
0.015 + 0.005 
-0.625 to +0.625 V 
0.015 + 0.008 
-0.3125 to +0.3125 V 
0.015 + 0.008 
-0.156 to +0.156 V 
0.020 + 0.008 
Notes: 
1) Specifications assume differential input single channel scan, 200 kHz scan rate, unfiltered. 
2) Accuracy specification is exclusive of noise.  Measurements were taken at P1. 
3) Inputs shorted to Signal Ground (SGND).  8192 samples.
 
 
Input Sequencer 
Analog, digital and counter inputs can be scanned synchronously based either on an internal programmable timer, or 
an external clock source.  Analog and digital outputs can be synchronized to either of these clocks. 
Scan Clock Sources: 2 
• 
Internal, programmable from 5 
µs to 5.96 hours maximum in 1 µs steps 
• 
External, TTL level input up to 200 kHz maximum 
Programmable Parameters per Scan: channel (random order), gain, unipolar or bipolar 
Depth: 16,384 locations 
On-board Channel-to-Channel Scan Rate: 5 or 10 
µs per channel, programmable 
Expansion Channel Scan Rate:  5 µs,10 
µs, or 1000 µs per channel, programmable 
External Acquisition Scan Clock Input   
Maximum Rate: 200 kHz 
Clock Signal Range: 0V to +5V 
Minimum Pulse Width: 50 ns high, 50 ns low 
External Sync Ports 
Number of External Sync Ports:  2, on rear panel 
Maximum Number of Units to be Synchronized:  4 units, scan-synchronous (post trigger) 
Maximum Length of Sync Cables: 15 feet (4.57 m), total for all cables 
Triggering 
Trigger Sources: 6, individually selectable for starting and stopping an acquisition. Stop acquisition can occur on 
a different channel than start acquisition; stop acquisition can be triggered via modes 2, 4, 5, or 6 described 
below.  Pre-trigger is supported with fixed or variable pre-trigger periods. 
1. Single-Channel Analog Hardware Trigger 
Any analog input channel, including any of the analog expansion channels, can be software programmed as 
the analog trigger channel. 
 
Input Signal Range: -10V to +10V max 
 Trigger 
Level: Programmable, 11-bit resolution 
 Hysteresis:  Programmable, 11-bit resolution 
 Latency:  5 µs max