Agilent Technologies Switch/Measure Manuale Utente

Pagina di 216
92
34980A User’s Guide
3
Introduction to the Plug-In Modules for the 34980A
Module Considerations
This section lists important items and actions that can affect the 
operation of your modules. 
General Considerations
Environmental Operating Conditions
These modules are designed to operate in a temperature range of 0 °C to 
+55 °C with non- condensing humidity. The maximum humidity is 80% at 
40 °C or higher. Do not use in locations where conductive dust or 
electrolytic salt dust may be present.
These modules should be operated in an indoor environment where 
temperature and humidity are controlled. Condensation can pose a 
potential shock hazard. Condensation can occur when the modules are 
moved from a cold to a warm environment, or if the temperature and/or 
humidity of the environment changes quickly.
The following table shows maximum voltage ratings for each module. 
If conditions change, ensure that condensation has evaporated and the 
instrument has thermally stabilized until pollution degree 1 conditions 
are restored before turning on power to the equipment. 
N O T E
To reduce wear on the internal DMM relays, wire like functions on 
adjacent channels. 
Module
Pollution Degree 1 Specifications
Pollution Degree 2 Specifications
34921A
40 channels, 300 V rms or DC, 1 A, 
60 VA per channel 
40 channels, 100V rms or DC, 1 A,
60 VA per channel
34922A
70 channels, 300 V rms or DC, 1 A 
60 VA per channel
70 channels, 100 V, 1 A, 
60 VA per channel
34923A
20/40/80 channels, 150 Vpeak, 0.5 A, 
10 VA per channel
20/40/80 channels 100 Vpeak, 0.5 A, 
10 VA per channel
34925A
40/80 channels, 80 Vpeak, 50 mA
40/80 channels, 80 Vpeak, 50 mA
34931A
Dual 4x8 matrix, 300 V rms or DC,
1 A, 60 VA per channel
Dual 4x8 matrix, 100 V rms or DC, 
1 A, 60 VA per channel
34932A
Dual 4x16 matrix, 300 V rms or DC,
1 A, 60 VA per channel
Dual 4x16 matrix, 100 V rms or DC, 
1 A, 60 VA per channel
34933A
Dual/quad 4x8 matrix, 150 Vpeak,
0.5 A, 10 VA per channel
Dual/quad 4x8 matrix, 100 Vpeak, 
0.5 A, 10 VA per channel