Pressure Systems, Inc.                                                                              98RK-1 & 9816 User’s Manual© 
 
 
 
Page 132 
 
www.PressureSystems.com 
5.1.6.9  
Supply Air and Purge Air Sensing Transducers 
 
The pneumatic input manifold (pneumatic backplane) on the 98RK-1 Scanner Interface Rack 
contains two (2) additional DH200 transducers to provide facility measurements for additional 
diagnostic capabilities.  The transducers read supply air pressure and purge air pressure.  
Proper supply air is required to move the calibration valve to implement such functions as re-
zero or purge.  Proper purge pressure is required to blow contaminants away from the module.  
Without a sufficient amount of purge pressure, contaminants could migrate back into the 
calibration valve assembly and cause catastrophic damage to the scanner. Pressures detected 
by these transducers may be read through the  NUSS application or through the TCP/IP 
connection with an appropriate ‘r’ command.  The pressures associated with these transducers 
are read as channels 17 and 18 (of a sixteen (16) channel scanner) and are displayed before 
the scanner channel pressures. 
 
Following are procedures for replacing the supply air and purge air sensing transducers and/or 
their associated O-rings on 98RK-1 racks. 
 
(1) 
Make sure the 98RK-1 Scanner Interface Rack power switch is in the OFF position and 
the unit is unplugged from electrical power.  Make sure there is no pressure applied to 
any of the input ports. 
 
 
(2) 
Remove the eight (8) Phillips-head screws (front and side) holding the backplane panel in 
place.  Gently pull the panel out and allow it to rest on the power supply leads and the 
ribbon cable connecting the rear panel to the back plane. 
 
Note
 
As you are looking at the pneumatic backplane, the two 
transducers are on the right side of the pneumatic input manifold; 
the purge-sensing transducer is on the right and the supply-
sensing transducer is on the left.
 
 
(3) 
Remove the two grommets that cover the access holes in the top of the 98RK-1 chassis 
(directly over each transducer).  Note: it may be easier, with less opportunity to abrade 
the chassis, to punch the grommets out from underneath, using your Allen-head or 
Phillips-head screwdriver. 
 
(4) 
Remove the 3/32" Allen-head screw from each of the transducers, being careful not to 
drop the screw behind the manifold assembly.  These screws attach to stand-offs and act 
as stabilizers for the flex circuit. 
 
 
(5) 
Unplug the eight-pin flex circuit from both DH200s (transducers).  The flex circuit is a 
single-piece unit that must be unplugged from both transducers. 
 
 
(6) 
Remove the transducer by removing the 4-40 threaded stand-off screw and lifting the 
transducer straight up from the mounting block.  Make sure the two (2) O-rings remain 
with the transducer as it is removed and the mounting block is clean and free of 
contaminants.  To replace the transducer O-rings, use the procedure as described in 
Section 5.1.4. If the entire transducer is to be replaced, make sure to also replace the O-
rings.  Additionally, ensure that the replaced transducer is the same pressure rating as 
the old one. 
 
Note
 
The pressure rating of the purge-sensing transducer is 750 psi and 
the pressure rating of the supply-sensing transducer is 150 psi.