2.3.4 Strain Gauge
The figure below shows the wiring hookup for 4-wire bridge input.
Figure 2.7
a) 4-Wire Voltage/Bridge Input
b) 4-Wire Bridge Input with
with Internal Excitation
External Excitation Wiring
Wiring Hookup
Hookup
In 4-Wire connections the voltage drop across long excitation lead wires of strain
gauge bridge may cause measurement errors. The output of a strain gauge
bridge also depends on the stability of excitation voltage. To correct for voltage
drop and changes in excitation voltage, 6-wire input configuration and ratio
measurement are used.
gauge bridge may cause measurement errors. The output of a strain gauge
bridge also depends on the stability of excitation voltage. To correct for voltage
drop and changes in excitation voltage, 6-wire input configuration and ratio
measurement are used.
In order for the Ratiometric to work properly, the External Excitation should
not drop below 4.6 Vdc.
not drop below 4.6 Vdc.
The figure below shows 6-wire hookup for 6-wire bridge input.
Figure 2.8
a) 6-Wire Bridge Input with
b) 6-Wire Bridge Input with
Internal Excitation and
External Excitation and
Ratio Measurement Wiring
Ratio Measurement Wiring
Hookup
Hookup
9