ENTER COMMUNICATION OPTION MENU:
Press  
a
1) Press 
a, if necessary, until CNFG prompt appears.
Press  
d
2) Display advances to 
INPT Input Menu.
Press  
3) Press 
a, if necessary, until display advances to COMM
Communication Options Menu. 
Press  
d
4) Display advances to 
C.PAR Communication Parameters
Submenu.
If Communication Option is not installed, the display shows
NONE and skips to the Color Display Menu.
COMMUNICATION PARAMETERS SUBMENU:
Allows the user to adjust Serial Communications Settings of the instrument.
When connecting an instrument to a computer or other device, the
Communications Parameters must match. Generally the default settings 
(as shown in Section 5) should be utilized.
Press  
d
5) Display advances to 
BAUD Baud Submenu. 
BAUD SUBMENU:
Press  
d
6) Display flashes previous selection for 
BAUD value.
Press  
b
7) Scroll through the available selections:
300_, 600_, 1200_,
2400_, 4800_, 9600_, 19.2K.
Press  
d
8) Display shows 
STRD stored message momentarily and then
advances to 
PRTY only, if it was changed, otherwise press a to
advance to 
PRTY Parity Submenu.
PARITY SUBMENU:
Press  
d
9) Display flashes previous selection for “Parity”.
Press  
b
10) Scroll through the available selections: NO, ODD, EVEN.
Press  
d
11) Display shows 
STRD stored message momentarily and then
advances to 
DATA only, if it was changed, otherwise press a to
advance to 
DATA Data Bit Submenu.
DATA BIT SUBMENU:
Press  
d
12) Display flashes previous selection for “Data Bit”.
Press  
b
13) Scroll through the available selections: 7-BIT, 8-BIT.
Press  
d
14) Display shows 
STRD stored message and then advances to
STOP only, if it was changed, otherwise press a to advance to
STOP Stop Bit Submenu.
51