3M #3960 Manuale Utente

Pagina di 58
3M™ Rapid Detection Reader 
Running Quality Control 
14 
Running IQC 
IQC runs automatically every time the 3M Rapid Detection Reader is turned on.  IQC can also be 
initiated manually or configured to run at predefined intervals.  IQC takes less than 1 minute to 
complete.   
When IQC is initiated, the appropriate tests are performed on each port in each attached TM.  
IQC verifies the power supply voltage, system memory, cartridge transport system, LED function 
and incubator function for heated assays.  During IQC the system also determines system power, 
verifies the system clock and verifies the Reader calibration.   
Should there be a test in progress in a port, IQC will begin for that port once the test cartridge is 
removed on assay completion.  Results of the IQC are stored for each cartridge port in the IQC 
results database. 
 
To run IQC manually: 
1.  From the Main Menu screen (Figure 4), touch Run IQC.  If 
prompted, enter the Operator ID and/or Password using either the 
touch screen or barcode scanner.  To accept the displayed 
information, touch Accept.  The IQC Status screen is displayed 
(Figure 13). 
NOTE:  Both the Operator ID and Password are case-sensitive.  
IQC is started on all idle ports and will automatically start as ports 
become free. 
2.  Touch OK to return to the Main Menu.  If IQC fails, the TM port 
LED illuminates red, the port icon turns red and the failed port can 
no longer be used. e.g.,  
 
Figure 13.  IQC Status 
 
 
To run IQC automatically: 
To run IQC automatically at predefined intervals, the Reader is configured using the RCU or the 
CM.   
To configure the Reader to run IQC at defined intervals through the RCU, refer to Editing Settings 
Using the Reader Configuration Utility (RCU) section.  
To configure the Reader to run IQC at defined intervals through the CM, refer to the Managing 
Settings section.