Maintenance 
36 
Maintenance 
 
 
WARNING!  
To reduce the risk associated with hazardous voltage: 
• 
Unplug the 3M Rapid Detection Reader before cleaning. 
• 
Do not disassemble the 3M Rapid Detection Reader or any of its related 
components and accessories.  The instrument contains no operator-serviceable 
components. 
• 
Do not immerse any of the 3M Rapid Detection Reader components in any liquid. 
• 
Do not use the 3M Rapid Detection Reader if there is apparent damage to a power 
cord or power supply. 
To reduce the risk associated with potentially infectious patient samples: 
• 
Do not spill specimen or sample fluids on any of the 3M Rapid Detection Reader 
components or on the outside of the Test Cartridge. 
• 
If a spill occurs, disinfect external surfaces only, using a soft cloth containing a 
solution of 0.5% bleach, 70% isopropyl alcohol or 70% ethanol. 
 
The 3M Rapid Detection Reader requires no regular maintenance or calibration during its 
intended life span, other than exterior cleaning as needed. 
Any liquids on the external surface of a Test Cartridge must be cleaned before inserting into the 
ports of the TM. 
Exterior Cleaning 
 
CAUTION.   
To reduce the risk of instrument or accessory damage:   
• 
Do not immerse the 3M Rapid Detection Reader or related devices in any liquid. 
• 
Do not spill liquids on the 3M Rapid Detection Reader and related accessory 
devices.   
• 
Do not expose the 3M Rapid Detection Reader to temperatures above 60°C 
(140°F). Do not disinfect or sterilize the 3M Rapid Detection Reader using devices 
such as autoclaves, ultrasonic cleaners, or pasteurizers. Use of such methods to 
disinfect/sterilize the 3M Rapid Detection Reader will cause permanent device 
damage and void the warranty. 
The external surfaces of the TM and CM can only be disinfected using a solution of 0.5% bleach, 
70% isopropyl alcohol or 70% ethanol applied to a soft cloth.   
 
CAUTION.   
Possible device damage:   
• 
DO NOT USE STRONG BLEACH (>0.5% solution) OR ACETONE!  Oxidants and 
solvents could damage the 3M Rapid Detection Reader casing and touch screen.   
• 
Clean external surfaces only.