2.4.5.1.1
Type (Upper Lower)
Allows you to define the limit you are editing as either an upper or lower limit. An
Upper
limit fails if the trace exceeds the limit. A
Lower
limit fails if the trace falls below the limit.
Key Path:
Display
,
Limits
,
Limit 1
Display
,
Limits
,
Limit 2
Dependencies/
Couplings:
If a margin has already been set for this limit line, and this key is used to
change the limit type, then the margin value is reset to 0 dB.
State Saved:
Not saved in instrument state. Survives power cycle and preset. The
limit-line data is saved in files.
Factory Preset
and *RST:
Not affected by preset.
Factory
Default:
Limits are off by default. Upper for
Limit 1
, then a limit line is created.
Lower for
Limit 2
, then a limit line is created.
History:
Added with firmware revision A.03.00
Remote Command:
:CALCulate:LLINe[1]|2:TYPE UPPer|LOWer
:CALCulate:LLINe[1]|2:TYPE?
Example:
:CALC:LLIN2:TYPE LOW
sets limit line 2 as a lower limit.
:CALC:LLIN1:TYPE?
responds with the limit line 1 limit type.
2.4.5.1.2
Limit Display
Turns limit-line display
On
or
Off
. Either
Limit
or
Margin
, as well as
Test
, must be turned on to
turn on a limit test.
Key Path:
Display
,
Limits
,
Limit 1
Display
,
Limits
,
Limit 2
State Saved:
Not saved in instrument state.
SCPI Status Bits/
OPC Dependencies: No OPC dependencies.
Factory Preset
and *RST:
Off
Factory
Default:
Off (when a limit line is created)
History:
Added with firmware revision A.03.00