Agilent Technologies E4445A Manuale Utente

Pagina di 664
Chapter 2
69
Instrument Functions: A - L
AMPLITUDE / Y Scale
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
Sets the Attenuation to 10 dB. The only valid suffix is dB. If no suffix is
sent, dB is assumed.
If a value >5 is entered, 10 is used.
If a value
≤5 is entered, 2 is used for PSA, or 5 is used for ESA.
2.0.13
Max Mixer Lvl
Enables you to set the relationship between the highest signal that can be displayed (the
reference level) and the input attenuation. The relationship applies whenever the
Attenuation
is set to
Auto
. The relationship is that the attenuation is given by reference
level minus the max mixer level. For example, as the reference level changes, the input
attenuator changes to ensure that a signal at the reference level does not exceed the
Max
Mixer Lvl
setting.
Key Path:
AMPLITUDE / Y Scale
State Saved:
Saved in Instrument State
Factory Preset: –10 dBm
Range:
–50 dBm to –10 dBm
Remote Command:
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <power>
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?
Example:
POW:MIX:RANG -15 dBm