Chapter 2
Instrument Functions: A - L
Auto Couple
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
VBW/RBW
ratio is set to 1.0
Span/RBW
ratio is set to Auto
Auto Sweep Time
is set to Normal
FFT & Sweep
is set to Auto: Best Dynamic Range
ADC Ranging
is set to Autorange
Marker Count, Gate Time
is set to Auto
NOTE
Marker Trace
and
Printer
have an
Auto
setting, but are not affected by
Auto All
.
Remote Command:
:COUPle ALL|NONE
:COUPle?
Remote Command Notes: sets all the functions to the manual (not coupled) mode. There
is no front-panel key equivalent to the COUPle NONE command.
ALL
puts all the functions into the auto coupled mode.
Example:
COUP ALL
2.1.2
FFT & Sweep
Selects the FFT vs. Sweep key functions.
Key Path:
Auto Couple
Remote Command:
[:SENSe]:SWEep:TYPE AUTO|FFT|SWEep
changes the sweep type to FFT or swept, or it
lets the analyzer automatically choose the type of analysis to use based on current
instrument settings.
[:SENSe]:SWEep:TYPE?
Example:
SWE:TYPE FFT
2.1.2.1
Auto: Best Dynamic Range
This function is automatically activated when
Auto All
is selected. Selecting
Auto: Best
Dynamic Range
tells the analyzer to choose between swept and FFT analysis, with a primary
goal of optimizing dynamic range. If the dynamic range is very close between swept and
FFT, then it chooses the faster one.
While
Zero Span
is selected, this key is greyed out. The status of the FFT & Swept selection
is saved when entering zero span and is restored when leaving zero span.
Key Path:
Auto Couple, FFT & Sweep
Saved State:
Saved in instrument state