HP (Hewlett-Packard) 8360 Manuale Utente

Pagina di 508
 Coupling 
Programming Codes
SCPI: 
FM:FILTer:HPASs 
 
<num> sets the AC bandwidth to 100 
 for any value   1 
and sets the AC bandwidth to 20 Hz for any value   1 
Analyzer: 
NONE
 
[MOD], also see “FM” and “Modulation”.
FM Coupling DC
Function Group 
Menu Map 
Description
This 
 (Option 002 only) lets you set the FM input to be
DC-coupled. Use DC coupling for modulation rates below 100 
In this mode, the phase-locked loop is de-activated. This means that
the synthesizer is operating as an open loop sweeper. The synthesizer
will not be phase locked, and therefore, be aware that the phase noise
and CW frequency accuracy specifications do not apply.
An asterisk next to the key label indicates that DC FM coupling is
selected. The factory preset default is AC coupling.
For synthesizers 
 Option 002, see FM On/Off DC.
Programming Codes
SCPI: 
FM:FILTer:HPASs <num>[freq 
Analyzer: 
NONE
 
 also see “FM” and “Modulation”.
 Operating and Programming Reference HP 8360
User’s Handbook