SECTION 2—INSTALLATION/SEQUENCE OF OPERATION
13
The electrical activation of the 4‐way Valve is accomplished every 8 to 15 seconds by the 
pressure sensor and P.C. Board electronics when the pressure reaches a set point of 21 
p.s.i. (144.79 kPa) output flows 4 L/min and above or 16 p.s.i. (110.32 kPa) output flows 3 
L/min and below. The time between cycles is dependent on altitude, flow rate and 
internal environmental factors.
A P.E. valve opens just prior to the shift of the 4‐way valve. This allows highly 
concentrated oxygen to enter the just exhausted bed from the top. This additional 
pressure allows the bed to start its cycle at a higher pressure. The P.E. valve will close just 
after the shift of the 4‐way valve.
If main power is lost, the Battery Free Power Loss Alarm will sound a short ʺBEEPʺ, with 
a long pause after. All units are equipped with a diagnostic alarm system that signals if 
the pneumatic pressure or electrical systems malfunction. The troubleshooting guide in 
this manual explains the alarm system signals and reasons, in detail. Refer to 
Troubleshooting on page 16.
SensO
2
 Oxygen Sensor Technology - Ceramic Zirconia 
Sensor
Technical Description
The oxygen being produced by the concentrator flows out of the product tank and into 
the flowmeter. A small flow of oxygen produced by the unit is sent through a precision 
orifice to the oxygen sensor mounted on the printed circuit board.
As the oxygen enters the sensor, it passes through a screen and contacts the sensing disk.
Electric current flowing through a metal film resistor heats the disk in excess of 300°
 
C. 
Oxygen molecules contact the electrode of the disk and pick‐up extra electrons to become 
oxygen ions. These oxygen ions are attracted to the electrode on the bottom of the zirconia 
sensing disk. Because of the crystal structure of the zirconia, only oxygen ions can pass 
through. When the oxygen ions reach the bottom electrode, the extra electrons are 
released from the oxygen ions and oxygen molecules return to the air. The number of 
electrons is directly related to the oxygen concentration. The electrons travel to the P.C. 
board where they are counted and the oxygen concentration reading is calculated.
A microprocessor on the P.C. board contains software that interprets the signal being 
received from the sensor. It compares the signal to clinically acceptable limits. Signals 
outside of the clinically acceptable limits generate responses in the form of lights, audible 
indicators, and/or system shut‐down.
Operating Sequence
Once the power switch has been turned on ( ), the SensO
2
 circuit will wait five minutes 
for the concentrator to begin producing clinically acceptable oxygen and the oxygen 
sensor to stabilize. The GREEN light will illuminate (indicating normal system operation) 
while the oxygen sensor is warming up.