SECTION 22—TIMING
75
Manual Tuning
NOTE: For this procedure, refer to FIGURE 22.2.
1. Turn power off ( ) and unplug the unit. 
3. Plug unit in but DO NOT turn on ( ) until instructed to below.
4. Locate the timing switch on the P.C. board.
5. Insert a small common or flathead screwdriver into the adjustment screw of the timing 
switch.
6. Set the timing switch initially at the number “7” position.
7. Turn unit on ( ) and wait 15 minutes for unit to stabilize.
8. If after 15 minutes of run time the concentration is below specification, change setting 
to a number one position lower than the previous setting.
9. Let unit run ten minutes. If concentration is lower still, change the switch to number 
two positions higher.
NOTE: You will find that the concentrator will perform better at either the higher or lower setting. 
The normal adjustment range is between timing positions 3 and A.
10. Continue to change setting one step at a time (up or down) until the concentration is 
the highest value achievable. 
11. Once concentration has reached specification, retiming is complete.
FIGURE 22.2   Manual Tuning
/O
I
I
Manual Timing  
Switch
P.C. Board
Control Panel