Chapter 4
Getting Started
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
4-21
Oxygen Source Connection Instructions
1. Locate the O
2
 filter with water trap (1) located in the lower left corner 
of the rear panel.
2. Ensure that the O
2
 filter and water trap (1) are properly attached to 
the ventilator. Inspect the bowl (1) for cracks or potential leaks. 
Ensure all connections are tight.
3. Inspect the O
2
 hose (2), hose connector (3) and the hose connector 
mate (4).
4. Check the oxygen gas supply. Clean, adjust pressure, and drain 
condensate from any water traps, filters, or regulators in the O
2
 supply 
lines.
5. Connect O
2
 gas connector (3) to the ventilator O
2
 filter with DISS hose 
connector (4).
6. If gas source is not already turned on, turn it on and verify that no O
2
 
gas is leaking.