Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Range parameters
↓↓
In the menu "Range parameters" and its submenus you can set up and control range
parameters like
parameters like
♦
begin and end of range
♦
response time (t
90
-time)
♦
autoranging
for one channel. You can change to the other channels of an MLT analyzer or analyzer
module with the F3 key.
module with the F3 key.
Parameters setup or changing to submenus:
♦
Select any line of variables or any menu line with the
↓↓
-key or the
↑↑
-key.
♦
Select the variable or change to the submenu with the
88
-key or the
→
→
-key.
♦
Select the parameter with the
↑↑
-key or the
↓↓
-key.
♦
Confirm the new setup with the
88
-key or
cancel and go back to the last setup with the F2 key.
Line of variables "Range control":
The parameter selected in this line is valid for the autoranging control of the current
channel (see p. 5-29). You have the following options:
channel (see p. 5-29). You have the following options:
Measure
Channel
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Range parameters --
Begin and end of ranges...
Autoranging control...
Range control:
Manual
Actual range number:
1
Response times (t90)...
5.1 Analyzer Module Setup