Emerson I-0021 Manuale Utente

Pagina di 238
5 - 32
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
• 
You can repeat the last two steps as often as necessary until you have selected all
interference components (maximum 3) you need.
4) Give the interference component into the sample gas channel and wait for a stable
signal.
5) Go to the line "Calculate factor for selected interference component !" with the 
↑↑
 -key or
the 
↓↓
 -key and then start the calculation with the 
88 
 -key or the 
 -key. If you want to
see the results change to the line "Interference factors..." and press the 
88 
 -key to go to
the corresponding submenu:
In this submenu you can also set up each interference component factor manual.
6) Select "Enabled" in the line "Compensation is" of the menu "Cross Interference
Compensation" to start the cross interference compensation.
Function line "Remove selected component !"
If you press the 
88 
 -key or the 
 -key in this line you will delete the interference factor of
the selected interference component !
Measure
Back...
MLT25/CH2/R2
F1
F2
F3
F4
F5
95.00 ppm
-- Channels --
MLT25/CH2
MLT25/CH3
MLT25/CH5
-
-
-
Measure
Back...
TAG
F1
F3
F4
95.00 ppm
1. Interference component factor:
1.000
1.000
2. Interference component factor: