Emerson I-0021 Manuale Utente

Pagina di 238
 3 
Startup and Operation, General Notes and Main Menu
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
3 - 3
3.4
Lines and Softkey Functionality
Lines can be selected by the 
↓↓
 -key or the 
↑↑
 -key. The selected line is displayed white on
black. You have four different types of lines in the menu:
Menu line... / Menu Softkey...
♦ 
Line/Softkey lettering ending with three dots.
♦ 
You will go to a submenu/further menu by pressing the softkey resp. by pressing the
88 
 -key or the 
 -key in the menu line.
Function line / Function Softkey !
♦ 
Line/Softkey lettering ending with an exclamation-mark.
♦ 
You will start a function (e.g. calibration) by pressing the softkey resp. by pressing the
88  
-key or the 
 -key in the function line.
Line of variables:
♦ 
Line ending with a colon.
♦ 
Display of module parameters (variables).
♦ 
Some parameters can be changed (e.g. begin of range), some parameters display only
a status (e.g. temperature) and cannot be changed. These variables will be displayed
below a line within the menu.
Text line
♦ 
Line without any punctuation marks.
♦ 
Only display of informations.
The following illustrations shall make plain the functionalities of lines and softkeys des-
cribed above:
Measure
Status...
Channel
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Analyzer Module Calibration --
Calibration procedure status...
Start span calibration procedure !
Expected span gas:
50.00  ppm
Span gas name:
Methane
Start zero calibration procedure !
Expected zero gas:
0.00  ppm
Valves...
Valve position:
Samplegas
Time & Date:
12:50:55  June 16, 1998
Menu Line
Function Lines
Lines of Variables
(These here cannot
be changed)