NGA-2000 System Calibration
Supplement - 12
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
A further general parameter is "Timeout for Gastest". It is located in the menu "Test
procedures".
SCTESTGAS
SCTIMEOUT
SCCALTIME1
SCSTEPMOD
This parameter is related to the possibility to activate a gas valve of a specific module for test
purposes. Here we can determine a time after which an activated gastest automatically
switches back onto SAMPLE-Gas.
Is this parameter set to "0" no automatic back switching is done.
System configuration and diagnostics...
System calibration...
Calibration/Test Procedures...
Test procedures...
-- Test Procedures --
Gastest of specific module                        SAMPLE-Gas
Timeout for Gastest (0 = no timeout)                      180 s
Choose specific analyzer module...
_______________________________________________
Procedure time:                                                           10 s
Analyzer-module:
                                 TFID