Emerson I-0021 Manuale Utente

Pagina di 238
4 - 38
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Measure
Channel
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Calibration Deviations --
Deviation from zero:
0.34  ppm
Deviation from span:
0.82  ppm
Sum of zero deviations:
4.41  ppm
Sum of span deviations:
7.57  ppm
Display
Status...
Main...
Channel
Calib...
TAG
F1
F2
F3
F4
F5
Failures:
No
Maintenance-Requests:
No
Temperature:
20.0 C
Operation:
Ready
Range: 1
37.50
ppm CH4
0.00
50.00
0.0
100.0
⇒ 
⇒ 
Change to the single
component display of
the current channel
Press  
F1
Options:
• 
With the F3 key you can
change to further available
channels to check their
"Calibration Deviations".
• 
With the F4 key you can
go back to the menu
"Calibration Procedure
Status".
⇒ 
⇒ 
Back to the single