4.11 Close all Valves
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 65
Measure
Status...
Channel
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Set Gas Valves --
Flow sample gas !
Flow zero gas !
Valve position:
AllClosed
Operation status:
Ready
Flow span gas !
Close all valves !
Flow test gas !
Display
Status...
Main...
Channel
Calib...
TAG
F1
F2
F3
F4
F5
Failures:
No
Maintenance-Requests:
No
Temperature:
20.0 C
Operation:
Ready
Range: 1
37.50
ppm CH4
0.00
50.00
0.0
100.0
⇒
⇒
Change to the single
component display
of the current
channel
component display
of the current
channel
Options:
•
Start the gasflow with
zero gas, span gas,
sample gas or test gas.
zero gas, span gas,
sample gas or test gas.
•
F3 key: Changing to
another available channel
to execute gas flow.
another available channel
to execute gas flow.
•
F4 key: Go back to the
menu "Analyzer Module
Calibration" to start
zeroing or spanning.
menu "Analyzer Module
Calibration" to start
zeroing or spanning.
⇒
⇒
Single component
display appears in
the standby mode:
All valves are closed
display appears in
the standby mode:
All valves are closed
Note:
After the valves are closed
you can start again the
measurement with the "Flow
sample gas" order (see 4.9).
After the valves are closed
you can start again the
measurement with the "Flow
sample gas" order (see 4.9).