Emerson I-0021 Manuale Utente

Pagina di 238
5 - 6
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Calibration parameters
↓↓
Span gases
↓↓
In the menu "Span Gas Definition" you can set up the default value for each range of the
current channel, the actual span gas concentration and the desired span gas unit. The
concentration of each range should be a value between 70 and 110% of the end of range
value.
Set up parameters:
♦ 
Select any line of variables with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable with the 
88 
 -key or the 
 -key.
♦ 
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key
resp. select the whole parameter with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the 
88 
 -key or
cancel and go back to the last value with the F2 key.
The last two lines of variables:
These lines are only information lines. The variables cannot be changed.
Note:
The "ppm 
 "mg/Nm
3
" conversion factor can be set up in the menu "General
Measure
Back...
MLT25/CH2/R2
F1
F2
F3
F4
F5
95.00 ppm
-- Span Gas Definition --
Actual span gas concentration:
250.0  ppm
Spangas range -2:
250.0  ppm
Spangas range -3:
2500  ppm
Spangas range -4:
10000  ppm
Span gas unit:
ppm
"ppm" --> "mg/Nm3" conversion factor:
1.000
Concentration in span gas units:
95.00  ppm
Spangas range -1:
50.00  ppm
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Span gases