Agilent Technologies 664xA Manuale Utente

Pagina di 67
Language Dictionary     33
*TRG
 
 Meaning and Type
 Trigger    Device Trigger
 
 Description
 This command generates a trigger when the trigger subsystem has BUS selected as its source.  The command has the same
affect as the Group Execute Trigger (<GET>) command.
 
 
Command Syntax
*TRG
 
Parameters
(None)
 
Query Syntax
(None)
 
Related Commands
ABOR    CURR:TRIG    INIT    TRIG[:IMM]   VOLT:TRIG
 
<GET>
 
*TST?
 
 Meaning and Type
 Test    Device Test
 
 Description
 This query causes the power supply to do a self-test and report any errors (see "Selftest Error Messages" in "Chapter 3 -
Turn-On Checkout" of the power supply Operating Guide).
 
 
Query Syntax
*TST?
 
Returned Parameters
<NR1>
 
0
Indicates power supply passed self-test.
 
Nonzero
Indicates an error code.
 
Related Commands
(None)
 
*WAI
 
 Meaning and Type
 Wait to Continue      Device Status
 
 Description
 This command instructs the power supply not to process any further commands until all pending operations are completed.
"Pending operations" are as defined under the *OPC command. *WAI can be aborted only by sending the power supply a
GPIB DCL (Device Clear) command.
 
 
Command Syntax
*WAI
 
Parameters
(None)
 
Query Syntax
(None)
 
Related Commands
*OPC   *OPC?