Agilent Technologies 664xA Manuale Utente

Pagina di 67
Language Dictionary     37
DISP:MODE
 
 Switches the display between its normal metering mode and a mode in which it displays text sent by the user.  The command
uses the character data <CRD> format.
 
 
Command Syntax
DISPlay[:WINDow]:MODE NORMalITEXT
 
Parameters
<CRD>    NORMal | TEXT
 
*RST Value
NORM
 
Examples
DISP:MODE NORM      DISPLAY:MODE NORMAL
 
DISPLAY:WINDOW:MODE TEXT
 
Query Syntax
DISPlay[:WINDow]:MODE?
 
Returned Parameters
<CRD>   NORMAL or TEXT
 
Related Commands
DISP    DISP:TEXT    *RST
DISP:TEXT
 
 Allows character strings to be sent to display.  The characters will be displayed when the display mode is TEXT.  The LCD
has the following character set:
 
 LCD Character Set
 uppercase letters
 A through Z (Case-sensitive entry)
 digits
 0 through 9
 punctuation
 _  |  “  $  < >  +  -  /  =  ?  .  :  ,
 blank space
 
 
 A display is capable of showing up to 12 characters.  However, the three punctuation characters do not count toward the 12-
character limit when they are preceded by an alphanumeric character.  When punctuation characters are included, then the
maximum number of characters (alphanumeric + punctuation) that can be displayed is 15.  If it exceeds the display capacity,
a message will be truncated to fit and no error message will be generated.  If any character in the message is not a member
of the above character set, the character will not be rejected but will be displayed as a "starburst" (all 16 segments of the
character will light).
 
 
Command Syntax 
DISPlay[:WINDow]:TEXT [:DATA] <STR>
 
Parameters 
(See LCD character set)
 
*RST Value
‘      ‘
 
Examples
DISP:TEXT "DEFAULT_MODE"
 
DISPLAY:WINDOW:TEXT:DATA    '533.2E-1VOLTS'
 
Query Syntax
DISPlay[:WINDow]:TEXT?
 
Returned Parameters
<STR>      (Last programmed text string)
 
Related Commands
DISP    DISP:MODE    *RST
 
Note
IEEE Standard Digital Interface for Programmable Instrumentation requires that a string be
enclosed in either single (‘) or double (") quotes.