Agilent Technologies 664xA Manuale Utente

Pagina di 67
Language Dictionary     41
Status Subsystem
 
 This subsystem programs the power supply status registers.  The power supply has three groups of status registers;
Operation, Questionable, and Standard Event.  The Standard Event group is programmed with Common commands as
described in "Chapter 4 - Status Reporting".  The Operation and Questionable status groups each consist of the Condition,
Enable, and Event registers and the NTR and PTR filters.
STAT:PRES
 
 This command sets all defined bits in the Status Subsystem PTR registers and clears all bits in the subsystem NTR and
Enable registers.  STAT:OPER:PTR is set to 1313 and STAT:QUES:PTR is set to 1555.
 
Command Syntax
STATus:PRESet
 
Parameters
(None)
 
Examples
STAT:PRES     STATUS:PRESET
 
Query Syntax
(None)
 
Related Commands
(None)
Status Operation Registers
 
 The bit configuration of all Status Operation registers is shown in the following table:
 
 Bit Configuration of Operation Registers
 Bit Position
 15-12
 11
 10
 9
 8
 7
 6
 5
 4
 3
 2
 1
 0
 Bit Name
 NU
 NU
 CC
 NU
 CV
 NU
 NU
 WTG  NU
 NU
 NU
 NU
 CAL
 Bit Weight
 
 2048
 1024
 512
 256
 128
 64
 32
 16
 8
 4
 2
 1
 CAL = Interface is computing new calibration constants; CC = The power module is in constant current mode.
 CV = The power module is in constant voltage mode; NU = (Not used); WTG = Interface is waiting for a trigger.
Note
See "Chapter 4 - Status Reporting" for more explanation of these registers.
STAT:OPER?
 
 This query returns the value of the Operation Event register.  The Event register is a read-only register which holds (latches)
all events that are passed by the Operation NTR and/or PTR filter.  Reading the Operation Event register clears it.
 
Query Syntax
 STATus:OPERtion[:EVENt]?
 
Parameters
(None)
 
Returned Parameters
<NR1>
(Register Value)
 
Examples
STAT:OPER?
STATUS:OPERATIOBAL:EVENT?
 
Related Commands 
*CLS     STAT:OPER:NTR     STAT:OPER:PTR
STAT:OPER:COND?
 
 This query returns the value of the Operation Condition register.  That is a read-only register which holds the real-time
(unlatched) operational status of the power supply.
 
Query Syntax
STATus:OPERation:CONDition?
 
Parameters
(None)
 
Examples
STAT:OPER:COND?     STATUS:OPERATION:CONDITION?
 
Returned Parameters
<NR1>      (Register value)
 
Related Commands
(None)