Agilent Technologies Model 6681A: 3250A-00101 Manuale Utente
Operation Verification 103
B
Operation Verification
Introduction
This appendix provides operation verification test procedures. The tests do not check all the operating parameters, but
verify that the power supply is performing properly. The required test equipment and acceptable test results are specified in
tables at the end of this appendix.
verify that the power supply is performing properly. The required test equipment and acceptable test results are specified in
tables at the end of this appendix.
Note
Performance Tests, which check all the specifications of the power supply, are given in the applicable
power supply Service Manual.
power supply Service Manual.
Test Equipment Required
List of Equipment
The following equipment is required to perform the tests:
Table B-1. Equipment Required for Verification Tests
Equipment
Characteristics
Recommended Model
Digital Voltmeter
Resolution: 10 nV @ 1 V
Readout: 8 1/2 digits
Accuracy: 20 ppm
Readout: 8 1/2 digits
Accuracy: 20 ppm
Agilent 3458A
Current Monitor Resistor
Agilent 6641A, 51A, 52A
100 A, 0.01
Ω
, 0.04%, 100 W
Guildline 9230/100
Agilent 6642A, 43A, 44A, 45A, 53A, 54A, 55A
15 A, 0.1
Ω
, 0.04%, 25 W
Guildline 9230/15
Agilent 6671A
300 A, 0.001
Ω
, 0.04%, 100 W
Guildline 9230/300
Agilent 6672A, 73A, 74A, 75A
l00 A, 0.001
Ω
, 0.04%, 100 W
Guildline 9230/100
Agilent 6680A, 81A
1000 A, 0.1 m
Ω
, 0.05%
Burster 1280S
Agilent 6682A, 83A, 84A
300 A, 0.1 m
Ω
, 0.05%
Guildline 9230/300
Current Monitoring Resistor
The 4-terminal current-monitoring resistor listed in Table B-1 is required to eliminate output current measurement error
caused by voltage drops in leads and connections. The specified resistors have special current-monitoring terminals inside
the load connection terminals. Connect the current monitor directly to these current-monitoring terminals.
caused by voltage drops in leads and connections. The specified resistors have special current-monitoring terminals inside
the load connection terminals. Connect the current monitor directly to these current-monitoring terminals.