Omega LVF-210 Manuale Utente

Pagina di 7
Step One
SPECIFICATIONS
1,600
1,400
1,200
1,000
800
600
400
200
0
12
18
24
30
36
Acceptable
Range
Unacceptable
Range
4-20 mA Sensor
Electrical Loading Limits
Max.
Series
Resistance
(Ohms)
Supply Voltage (VDC)
100
80
60
40
20
00
-20
12
16
20
24
28
32
36
Acceptable
Range
Unacceptable
Range
Maximum Tem. / Voltage Derating
Continuous 20 mA Sinking Curve
Ambient
Sensor
Temperature
(¡C)
Operating Voltage (VDC)
Orientation:
Accuracy:
Repeatability:
Supply voltage:
Consumption:
Contact type:
Contact rating:
Contact output:
Process temp.:
Pressure:
Sensor rating:
Cable type:
 
Cable length:
Process mount:
Mount. gasket:
Classification:
CE compliance:
Universal
±
 1 mm in water
±
 0.5 mm in water
12-36 VDC
25 mA maximum
(1) SPST relay
GP: 120 VAC/VDC @ 1A
(CE: 60 VAC/VDC @ 1A)
Selectable NO/NC
F: -40˚ to 194˚ 
C: -40˚ to 90˚
150 psi (10 bar) @ 25˚ 
C., derated @ 1.667 
psi (.113 bar) per ˚C.
above 25˚ C.
NEMA 6 (IP68)
4-conductor, #22 AWG
(shielded)
Standard: 10' (3m)
3/4" NPT (3/4" G / Rp)
Viton® (G version only)
General purpose 
EN 50082-2 immunity
EN 55011 emission
EN 61010-1 safety
Sensor material:
Cable jacket mat.:
PP-Ryton® (glass fill)
PP 
Sensor material:
Cable jacket mat.:
-150/-152: PP
-151/-153: PFA
-150/-152: PP
-151/-153: PFA
Dielectric range:
Conductive range:
Sensor material:
Cable jacket mat.:
> 20 constants
> 100 micromhos
PP
PP 
Dielectric range:
Conductive range:
Sensor material:
Cable jacket mat.:
> 20 constants
> 100 micromhos
-101/-103: PP
-102/-104: PFA
-101/-103: PP
-102/-104: PFA
Sensor material:
Cable jacket mat.:
-210/-212: PP
-211/-213: PFA
-210/-212: PP
-211/-213: PFA
Common Specifications:
LVC-152 series Specifications:
LTU-101 series Specifications:
LVU-150 series Specifications:
LVF-210 series Specifications:
LVC-100 series Specifications:
160
140
120
100
80
60
40
20
00
-40 -20
00
20
40
60
80
100
Acceptable
Range
Unacceptable
Range
Pressure / Temperature Derating
Operating
Pressure
(psi)
Temperature (¡C)
Acetone
21
Acetoaldehyde
22.2
Acetyl methyl hexyl
ketone
28
Alcohol
16 to 31
Ammonia
15 to 25
Acetic acid
4.1 to 6.2
Butyl chloride
9.6
Barium chloride 9 to 11
Benzene
2.3
Benzine
2.3
Barium nitrate
5.6
Bromine
3.1
Chlorobenzene 4.7 to 6
Chlorotoluene
4.7
Chloroform
4.5 to 5.0
Chlorine, liquid
2.0
Carbon tetrachloride 2.2
Cyan
2.6
Cyclohexane methanol
3.7
D.I. Water
20
Ethyl toluene
2.2
Ethyl alcohol
23
Ethylene glycol
37
Ethylene oxide
14
Ethylene dichloride
11 to 17
Ethylene chloride
10.5
Ethyl acetate
6.4
Ethyl salicylate
8.6
Ethyl stearate
2.9
Ethyl silicote
4.1
Formic acid
59
Ferric oleate
2.6
Freon
2.2
Glycerine
47
Glycol
30
Glycol nitrite
27
Gasoline
2 to 2.2
Hydrochloric acid
4.6
Isobutyric acid
2.7
Isobutyl methyl ketone
13
Jet fuel
1.7
Lead carbonate
18
Lead nitrate
38
Methyl salicylate
9.0
Methanol
33
Methyl alcohol 33 to 38
Margarine, liquid
2.8 to 3.2
Methyl acetate
7.3
N-butyl formate
2.4
Nitrobenzene
26 to 35
Nitrotoluene
25
Naphthalene
2.3 to 2.5
Oils, vegetable2.5 to 3.5
Oils, mineral 2.3 to 2.4
Oils, petroleum
1.8 to 2.2
Oleic acid
2.5
Propane, liquid
1.8 to 1.9
Potassium nitrate
5.0 to 5.9
Potassium chloride
5.0
Stearic acid
2.3
Toluene
2.4
Trichloroethylene
3.4
Trichloroacetic acid 4.5
Terephthalic acid
1.5 to 1.7
Thinner
3.7
Urea
3.5
Vinyl chloride
2.8 to 6
Vinyl alcohol 1.8 to 2.0
Water, 20°C
80
Water, 100°C
48
Table of Common Dielectric Constants
NOTE: Liquids with a dielectric constant less than 20 will not be detected by an LVC-152 series or LVC-100 series level switch, as