Printronix SL4M Manuale Utente

Pagina di 272
47
Running Auto Calibrate
You can initiate Run Auto-Cal (Auto Calibrate) via the the SENSOR SETUP 
menu.
NOTE: Verify that the Gap/Mark Sensor option (Gap, Mark, or Disable) 
Check that the media sensors are horizontally positioned to permit 
sensing of the label length indicators. See “Positioning The Media 
Sensors” on page 44.
If Use Label Length is enabled in the SENSOR SETUP menu, make 
sure the Label Length value entered in the MEDIA SETUP menu 
matches the physical length of the installed media. Entering the 
correct length forces the printer to advance media far enough during 
calibrate for long labels (so actual gaps and marks can be detected).
If you try to do an Auto Calibrate when Peel-Off Media Handling is 
enabled, the LCD will display “CANNOT CALIBRATE/Disable Peel-
Off.” Before you can do an Auto Calibrate, you must select another 
media handling mode or enable the Cal in Peel mode option under 
the SENSOR SETUP menu. Cal in Peel Mode displays only if Admin 
User is Enabled under the PRINTER SETUP menu.
1.
Press 
 to place the printer in Menu mode.
2.
Press Down 
 (Enter) keys together until “THE 
↵ 
KEY IS UNLOCKED” 
displays.
3.
Press the Right key until SENSOR SETUP displays.
4.
Press 
 to enter the SENSOR SETUP menu.
5.
Press Down key until Run Auto-Cal displays.
6.
Press 
 to select the option. Media advances until it can accurately detect 
the label length indicators and then stops at the Top-of-Form position. 
The Sensed Distance value will then display for one second.
7.
Auto Calibrate is successful when the Sensed Distance value correctly 
matches that of the installed media:
Gap/Mark Sensor = Gap: The Sensed Distance value is the physical 
length of one label plus the length of one gap, notch, or hole.
Gap/Mark Sensor = Mark: The Sensed Distance value is the 
physical distance from the trailing edge of one black mark to the 
trailing edge of the next.
Gap/Mark Sensor = Disable: Not applicable. If Gap/Mark Sensor is 
set to Disable, the Sensed Distance value will not be updated.
If “GAP NOT DETECTED” displays, run auto calibrate again.
If auto calibrate continues to end with an incorrect Sensed Distance value 
displayed or a fault message displayed, run Manual Calibrate as 
described on page 50 or see Table 14 on page 184.
..
.