MaxiScan 3300 – Installation Guide – Edition 2.0
 3 Connections
RS-485 pinout
MaxiScan 15-pin subD female connector
I/O sync max current 20 mA
RS-485: connect pins 1 and 2 and connect
pins 3 and 4. Install resistor = user cable
characteristic impedance between pins 1 and 3