3  Measuring Two-point Distance or Entering Information—TOOL Menu 
- 72 - 
8) Press    to return to the previous menu. Any calibration values will be erased unless 
they have been saved. 
NOTE:  If you select "E" on the calibration screen, calibration values are converted 
to inches or millimeters before saving. Values entered in centimeters or 
millimeters are converted to inches, while values entered in micrometers or 
nanometers are converted to millimeters. 
 
< When using the magnification of the objective and relay lenses > 
 
1) 
Select a number in which you want to save the calibration value from    s1    through    s7 . 
Drag the mouse pointer over one of numbers   s1  through  s7 , and the calibration 
information for that number will be displayed in the Info area for your confirmation. 
NOTE:  If the microscope connected to your system is the Nikon 80i/90i microscope 
with motorized or intelligent nosepiece, note that each calibration value call 
number is linked to the hole number of the nosepiece. This means you 
cannot change calibration values by selecting    s1    through    s7 . The value 
is selected automatically from    s1    through    s7    based on the hole 
number of the nosepiece you are currently using. 
2) 
Press  Calib.  to bring up the calibration screen shown below. 
 
 
 
3) Click 
the 
  numeric value input box to display the numeric keypad. Use the 
keypad to enter the magnification of the objective. 
4) Click 
the 
  numeric value input box to display the numeric keypad. Use the 
keypad to enter the magnification of the relay lens. 
5) 
If you want to attach information to the calibration value you've created, click the Info 
input box and enter text for the information you want to attach. This information will 
prove helpful because you can confirm Info when you call up the calibration value at a 
later time. 
6) 
Check the Set number to confirm that the calibration number is correct, then click    Save 
to save it.    Save    is disabled if no value is entered. 
7) Press    to return to the previous menu. Any calibration values not saved are deleted. 
NOTE: 
If the microscope connected to your system is the Nikon 80i/90i microscope 
with motorized or intelligent nosepiece and the calibration method you 
selected is "using the magnification of the objective and relay lenses," you 
do not need to enter the magnification of the objective. If the camera head 
is connected to the rear port of the DIH, you do not need to enter the 
magnification of the relay lens. If you calibrated with one objective, other 
connected objectives will be also calibrated automatically with their 
magnifications being converted. 
If you selected the "using the magnification of the objective and relay 
lenses" calibration method, keep in mind that the calibration value may 
incorporate a certain amount of error attributable to optical errors of the 
objective. With the Nikon 80i/90i microscope connected, errors may range 
up to 10%.