6 - 2
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Structure of chapter six:
The structure of chapter six is analogous to the structure of chapter five:
At the begin of any chapter you will find the way to a certain submenu of the software
starting from the line "Expert controls and set up" in the "Main Menu". The way will be
described with the software catchwords that you have to enter one after another to reach
the corresponding submenu. At the end of the catchword listing you will find the illustration
of the LCD screen. After that you can read the set up instructions and explanations
containing further menu pictures, if necessary.
Example:
You want to optimize the converter efficiency.
Main Menu — Technical level configuration
↓↓
Diagnostic menus
↓↓
Analyzer module diagnostics
↓↓
NOx converter efficiency
↓↓
In this menu you can optimize NOx converter efficiency.
1) Set up the desired converter temperature (set point):
• 
Press the ENTER key or the 
 -key to select the parameter
• 
Select any digit you want to change with the 
 -key or the 
 -key and ...
Further explanations and instructions to set up the parameters will follow !
HOME
ESCAPE
NO/NOx
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
NOx converter efficiency
Converter temperature:
362.0 C
  Capillary pressure:
340.0 hPa
Measurement mode:
NOx
Measurement:
95.00 ppm
Last temperature reading:
361.7 C
Converter set point:
360.0 C
Last measurement (other mode):
0.000 ppm